Principles of Dialysis
Indications & Contraindications
Hemodialysis
Hemodialysis related therapies
Peritoneal Dialysis
Other
Minitips

Hemdialysis Machine

Contents:

Blood pump

Dialysate delivery system

Safety monitors      

Options

System disinfection:

Alarm during Hemodialysis

 

Hemodialysis machine consists of  : 

1.Blood pump

2.Dialysate delivery system

3.Safety monitors      

4.Options

5.System disinfection:

 

1. Blood pump

Function : It circulate the blood through the dialyzer back to the patient.

Kinds    : Presently peristaltic roller pumps which works by progressive compressing special segments of the blood tubing are used.

Pump occlusion: Occlusion means that the roller  compress the tubing segments against the semicircular housing sufficiently to close the lumen completely at that point. Consistent pump flow rates require precise occlusion. Overocclusion may crack the tubing and rupture of the pumping segment. If occlusion is incomplete, there will be backflow of the blood with each stroke, inaccurate flow rate and red cell damage The flow rate for adult is ~ 4 times patient body weight ml/minute, up to 600ml/minute in high efficiency and high-flux dialysis.

 

2.Dialysate delivery system :

Function

-Appropriate blending of concentrate and water for preparation of final dialysate.

-To monitor dialysate for temperature, composition, and blood leak

-To control dialysate pressure or ultrafiltration rate.

-To regulate the dialysate flow rate through the dialyzer.

- Deareation of water

-Provide protective mechanisms to isolate the blood circuit and the patient from unsafe dialysis.

-System for disinfection and cleaning

Types  a. Single batch system    b.Continuous proportioning system     c.Sorbent regenerative system

a. Single batch system     (is of historic interest)

b.Continuous proportioning system (early1960s)

1.Central      :

-All the dialysis solution used for dialysis is produced by mixing the concentrate with purified water and pumped to each machine

-Advantage : Cost effective

-Disadvantage :Not permitting dialysate composition modification

2.Individual    : 

-Each dialysis machine proportions its own concentrate with the purified water.

-Advantage :

1. Permitting dialysate composition modification by selecting different concentrate or altering the mixing ratio.

2. There is significant saving in time.

3. Less bacteriologic and chemical contamination. 4. Reduced the risk of human error.

-Disadvantages :

a. Complex sophisticated system  

b. Costly 

c.Troubleshouting may be difficult and needs a factory-trained service technician.

Methods of mixing in proportioning system 

1. Matched pumps 

2. Matched metering valves

3. Matched hydraulic-accumulation cylinders  

4. Electronic feedback circuit. Details of mixing are beyond the scope the Tips

c.Sorbent regenerative system  for renewing dialysate. At present it is rarely used.

 

Heating  

-The physiological range of dialysate temperature is 36-42 c

-Temperature below 36 patients will complain of cold and uncomfortable.

-Temperature above 42 protein denature can occur and above 45 red blood cells will hemolyze.  cardiorespiratory arrest and death (from hyperkalaemia and hypoxemia )as a consequence of overheated may occur. 

Deaeration or Degassing  

Water contains considerable amount of dissolved air and microbubbles. Once negative pressure is applied and water is warmed the air comes out of the solution as expanding microbubbles. 

Bubbles causes:

1.Artifacts on conductivity, temperature sensors and flow meters

2.When bubbles cross the membrane into the blood foaming occur and increases potential for clot formation, hemolysis and occlusion of the fibers.

3. It decreases the surface area of the dialyzer

4. It block the dialysate flow channels subsequently dialysate pressure drops and the dialysate surface area incompletely utilized for solute transport

Deareation devices uses warmers along with negative pressure (- 600 mmHg) to bring the dissolved air out of the solution.  An air trap or coalescing filter then capture the air and vents them out. 

Dialysate pressures

The dialysate delivery system must be capable of generating positive and negative dialysate pressures (~ - 400 to + 200).

·         In dialysis machines without an ultrafiltration controller the dialysate pump is located at the line leading from the dialyzer to the drain. This allows the machine to create a negative pressure in the dialysate compartment of the dialyzer. The negative pressure is generated by partial occlusion of dialysate hose proximal to the dialyzer. To increase the rate of ultrafiltration, negative dialysate pressures (suction pressures) are often required. Transmembranous pressure is a resultant of the blood compartment and dialysate compartment pressures.

·        High positive pressure is required in situation in which ultrfiltration coefficient or the blood compartment pressure are high to limits the rate of obligatory ultrafiltration. Higher positive pressure (+ 350) may be necessary with high-flux membranes However, backfiltration occurs when dialysate pressure is greater than that of blood compartment, e.g. at the blood outlet. Transport of endotoxin fragments into the blood in such membranes.

Dialysate Flow

The standard dialysate flow rate is 500 ml/minute.  When high-efficiency and high-flux  dialyzer are used high dialysate flow rate between 700 and 1000 is required along with high blood flow rate.   High flow rates are needed  for:

1. Optimum use of dialyzer surface area for solute transport. 

2.With high dialysate flow rate a positive dialysate pressure is generated which limits the rate of obligatory ultrafiltration.

 

3. Safety monitors:

Monitors for blood circuit    

1.Arterial pressure monitor 

Locations:  Proximal to the blood pump      

Function:  1. It reads the arterial pressure at the segment between the patient's needle site             and proximal to the blood pump which represent the negative pressure created by the roller pump. 2.It identifies how much suction is being placed on the arterial wall and guard against excessive suction on the vascular access, e.g. .if suddenly the arterial pressure increases from 100 to 200 mmHg this could indicate clotted or dislodged needle, low patient BP or kink in the arterial line. 3. Resistance within the needle (function of the gauge and needle length). 4. It provide an index of vascular access blood supply relative to the flow demand by the blood pump. 5. A guide to appropriatence of needle placement or kink or obstruction in the blood segment between the patient and the monitor.    

Mechanism: The pressure is  monitored by mechanical or electronic manometers (pressure transducers) . Electronic transducer is more sensitive and have rapid response

2.Venous pressure monitor 

Location: Just distal to the dialyzer, usually attached to the top of the venous air trap.

Function: 1. It reads the venous pressure at the segment between the point after the dialyzer and before the blood reenters the patient's body. 2. It represent the resistance of the blood returning to the patient via the venous needle e.g. if suddenly the venous pressure increases from 50 to 150 mmHg this could indicate kink in the venous line, clotted air trap or the venous  needle may be clotted or maligned. Sudden dropping venous pressure occurs when the venous needle is pulled out, wet transducer or clotted arterial chamber.3. It may indicate venous stenosis proximal to the needle site.  

Mechanism:  As in arterial pressure monitor.

3.Venous air trap and air detector

Location: 1. Just distal to venous pressure monitor. 

                2. Often second air trap on the arterial line is also used.

Function: To prevent air entry into the patient or to the dialyzer. It is used also to measure the pressure in that segment of blood circuit.

Mechanism: When air bubbles entered the blood circuit a sensor reacted through ultrasonic transducer or light beam by stopping the blood pump, clamping the venous line, and activating audiovisual alarms.

 

Monitors for dialysis solution circuit

1.Conductivity

Location :   Before dialysate reaches the dialyzer

Function    To guard against excessive diluted or concentrated dialysis solution.

Mechanism : When conductivity meter (conductivity probe) detects high or low conductivity  the machine automatically sounds an alarm and puts the dialysate into bypass mode so that  no dialysate flow to the dialyzer.

Exposure of the blood to hyperosmolar dialysate can lead to hypenatraemia and other electrolyte disturbances. Exposure of the blood to hypoosmolar dialysate can results in  hyponatraemia and rapid hemolysis.

 

2.Temperature

Location:    Before the dialyzer

Function:    To avoid high temperature

Mechanism: Through a temperature sensor, high temperature activate an audiovisual alarm simultaneously with bypass mechanism.

3.Bypass  valve or mechanism

If the conductivity or Temperature found to be out of limits a bypass valve is activated to divert dialysate around the dialyzer directly to the drain.

4.Blood leak detector

Location : In the effluent dialysate line

Function: It guards against undetected blood loss during dialysis. The maximum limit to hemoglobin present in dialysate is 70 mg/L (corresponding to blood loss of ~0.4 ml/minute ) after which the blood pump is stopped.

Mechanism: A beam of light is directed through a column of dialysate onto a photoelectric cell.  A change in translucence and light scatter in dialysate reduces the light received by the photocell, stopping blood pump and activating audiovisual alarms.

5.Dialysate pressure or transmembranous monitor

Function : To monitor the ultrafiltration with an upper limits avoids excessive level (the usual preset range 0 to -500 mmHg). Excessive TMP may lead to rupture of the dialyzer and secondary deareation (cause air accumulation).

Mechanism: The monitor may have automatic or manually set limits, so that extrusion outside the limits trigger an audiovisual alarm.

4. Options in hemodialysis machine:

1.Heparin pump

2.Bicarbonate  3.Variable sodium  

4.Controlled Ultrafiltration 

5.Programable ultrafiltration

6 Dialysate urea sensor [on-line Kt/V monitor] 7.Single blood pathway.

5. System disinfection:

All dialysis unit s must have written policies the deal with the dialysis fluid pathway and dialysis machine . Disinfection procedure should be done on regular base according to manufacturer's instructions.

Target: To control bacterial contamination. HIV, HCV and HBV viruses are known to be inactivated by common household bleach.

Methods

1.Heat disinfection requires temperature greater than 85-90 C

2.Chemicals disinfection such as formaldehyde ,  sodium hydrochloride and acetic acid

Machine surfaces, patient's chairs, surrounding furniture, equipment should be routinely wiped with 10% bleach solution following every patient dialysis. Blood spills should be cleaned immediately.  Leak proof bags should be used to transport linen soiled with blood or body fluid

 

6. Alarm during Hemodialysis- Problem Solving Guide Lines]

When an alarm is activated during dialysis do the following:

1-Identify which alarm has been activated.         

2-Identify the cause.

3-correct the cause                                            

 4-Resume dialysis if safe to do so

Problem

Management

1.Power

 

 

a. Turn the system off and on; if still no power, check the power cord; make sure power is available

b. Check the fuse

2. Arterial and venous Pressure

 

a. Check to see that blood pump is running and connected properly

b. Check to see if blood flow rate has changed

c. Determine if patient has coughed or moved

d. Check to see  if the monitor line is leaking

e. Check the blood line for kinks or leaks

f. Ensure the monitoring lines are connected to  proper drip chambers

 

2.1 High venous pressure

 

1-Manipulate the needle and or the line. If the access is small, a tourniquet must be used, being certain the blood pump is off ..Recantation with new needle if needed ,the original one should be left in place until the end of dialysis. to avoid undue bleeding as the patient is heparinized

 2-Adjust the blood flow rate , Proper heparnization , Treat access problem ,and Proper needle and needling

 3-Extreme care must be exercised when dialyzing a patient with high venous pressure

-This increases the baseline TMP and obligatory  ultrafiltration will occur

-Single-needle device is occasionally impossible to use, because with high  venous pressure ,venous return will be impaired and blood recirculation will be high

 

2.2 Increase  in negative pressure: (Excessive inflow suction)

Management of increased negative pressure:  

    1- Manipulation of the arterial needle is similar to that of venous needle. In most cases the needle may have to be replaced with the same precaution as with venous needle

    2-Treat the cause. 

   3- Proper needling   

   4-Asses the access

3.Air detector alarm

 

a-Check for air leaks around tubing joints.

b-Excessive undetected negative pressure

c-Check for unattended intravenous solution administration.

 Management of air embolism if present

a-Clamp the venous line and stop the blood pump.

b-Place the patient in Trendelenberg position on the left side with the

chest and head tilted downward to trap the air at the apex of right ventricle away from the outflow tract.

c-Cardiorespiratory support ,Oxygen 100%.

d-Occasionally percutaneous aspiration of the air foam from the heart may be necessary.

e-Other measures include IV Dexamthsone to reduce brain edema ,Heparin /Dextran to improve the microcirculation.

 

 

4.High  Positive dialysate pressure alarm

 

a Check to see if drain is occluded or kinked

b. Check to see if dialysate hoses are leaking air

5.Low Negative dialysate pressure alarm

Check to see if dialysate hoses are kinked

     

6.HighTemperature alarm

a. Determine temperature of dialysate (to dialyzer) in the line actually high

 b. Check to ensure incoming water temperature is blew 90 F

7.Low Temperature alarm

a. Determine temperature of dialysate”to dialyzer” in  the line  is actually Low

b. Turn the mode selector switch to “dialyze”

c. Allow adequate time for the system to stabilize and come to proper temperature range

d. Check to ensure incoming water temperature is above 40 F

8.High conductivity alarm

a. Check to see if water is flowing too slowly or turn off

 b. Check for kink in the concentrate out line

c. Make sure that the system  has had time to stabilize

d. Analyze the dialysate to confirm high conductivity at the “to dialyzer” line connection :

1.If normal, there is malfunction in the machine itself (change it)

2.If high again concentrate should be changed

3.Check the conductivity before resuming dialysis   

e. Be certain that the dialysate flow rate is proper.

 

 

 

 

9.Low Conductivity alarm

 

 

a. Turn the mode selector switch to “dialyze”

b. Connect the concentrate line to the system

c. Drop the concentrate line into the concentrate container

d. Check for kinks in the “concentrate” in line

e. Make sure the filter on the “concentrate in” is clean

g. Allow adequate time for the system to stabilize

h. Change the concentrate container if it is dry

i. If the concentrate container has not run dry ,a sample of dialysate should sent to the laboratory for sodium and chloride                

j. Analyze the dialysate to confirm low conductivity at drain

 1.If low conductivity is confirmed  ; change the concentrate bottle (recheck again)

 2.If the conductivity still low after changing the concentrate , a different machine should be  tried

l. Be certain that the dialysate flow rate is proper

10.Blood Leak alarm

a. Make sure the blood leak detector is clean

b. Check the effluent for traces of blood

c .Check the dialysate lines for gross blood leak

d. Exclude 1.bilirubin in dialysate in jaundiced patient

                2.air bubbles in dialysate             3.dirty sensor

e. If a leak confirmed and you can not mange the cause

   1.reduce dialysate compartment pressure to -50 mmHg to avoid bacterial entery to the blood

   2.Dicontinue dialysis

N.B. Review the previous tips and the manual for each machine if needed.